Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
4-44
A5 As A
13
A
12
A
11
A
23
A
22
A
21
A
33
A
32
A
31
a.
NORM field size: Set Field size to NORM, adjust trimmer A31.
b.
MAG1 filed size: Set Field size to MAG1, adjust trimmer A32.
c.
MAG2 filed size: Set Field size to MAG2, adjust trimmer A33.
3.
Take an exposure, and check the Field size, repeat as many times as necessary as field size is in
specification.
Cleanup
1.
After passing, remove the BAFT/UBAT from the I.I. Grid.
2.
Make sure to re-assembly of all components as the following
a.
Power off the system.
b.
Disconnect Camera Cables.
c.
Reinstall the I.I. Power Supply and connect cables between power supply and I.I.
Center Resolution
Edge Resolution
Field Size
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...