Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
3-56
5.
Connect video cable (PN: 5085792) of Video distributor to DVI output of computer.
6.
Remove cable (PN: 5085729), connect power cable PN: 5085793 to
computer power out, reconnect connector of SIB/PDU.
7.
Functional check: Power on the system. Check that the left
monitor will display GE logo screen. Disconnect and pull out
the left monitor video cable then connect to 2 OUTPUTs
respectively. Verify that the left monitor will display GE logo
screen in each case.
8.
Reconnect the video cable of left monitor and install the
cabinet and other covers of workstation. Install the plastic
protective cover of the external DVI OUTPUTs and Video
output.
视频输入口
视频输入口
视频输入口
DVI Output
DVI Input
Power