Calibration
4-53
Camera of Brivo OEC 865
5491851 ASM, GigE Camera GE1020B
Camera of Brivo OEC 865
6491850 ASM, GigE Camera GE1020C
The secure screw of 5085003 is 1/16 inch HEX.
Image Intensifier Focus Calibration
I.I. Focus adjustment is done in the I.I. factory. Adjust with much care only if the power
supply is changed or repaired. Always use non-metal flathead screwdriver to adjust the
I.I. potentiometer to avoid short circuit.
For Brivo OEC 715/ 785(I.I PN: 5075001) and Brivo OEC 865(I.I PN: 5075008):
1.
Loosen 2 securing screws for I.I. Power Supply.
Camera Focus
Adjustment Ring
Camera Focus
Secure
Screw
Camera Focus
Adjustment
Ring
Camera Focus
secure Screw