Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-62
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0401500
Error
Allocate
Memory failed
System run out of memory.
Please restart the system. If
this issue persists, please call
for service.
Re install software
0401501
Warning
Write PIPE
failed
Communication exception.
Please restart the system. If
this issue persists, please call
for service.
Re install software
0401502
Warning
Read PIPE failed Communication exception.
Please restart the system. If
this issue persists, please call
for service.
Re install software
0401503
Error
Set PIPE failed
Communication ini exception.
Please restart the system. If
this issue persists, please call
for service.
Re install software
0401504
Error
CORBA
communication
failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
Re install software
0401505
Error
Sync operation
failed
Execution exception. Please
restart the system. If this issue
persists, please call for service.
Re install software
0401506
Error
Thread
operation failed
Execution thread exception.
Please restart the system. If
this issue persists, please call
for service.
Re install software
0401001
Error
Open serial port
failed
Workstation and mainframe
communication exception.
Please restart the system. If
this issue persists, please call
for service.
0401002
Warning
Write serial port
failed
Workstation and mainframe
communication exception.
Please restart the system. If
this issue persists, please call
for service.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...