Calibration
4-39
Step 5 Calibrate Iris position between 60mm and 65mm:
Step 5 Calibrate Iris position into 50mm*50mm square:
Step 7 Calibrate shutter maximum position for Normal Mode:
(Move shutter untill the leaves just out
of the circular mask)
Comments: Measure method is same as Iris diameter measure method.
Diameter=60~65mm
50mm×50mm Square of BAFT