Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
7-36
5.
Refer to calibration matrix in the appendix to perform calibration and checks.
After calibration and check, shut down system and reinstall the PCB Box and mainframe back covers.
Mother board (P/N: 5085601)
1.
Remove the rear cover of the C-Arm console.
2.
Release the screws of e-Box cover and remove it.
3.
Remove the collimator control board, mainframe control
board and Filament Drive Board.
4.
Release the screws and remove the whole e-Box.
5.
Unplug all cables connected to the motherboard.
6.
Loose the screws and remove the mother board.
7.
Remove 3 PWA insert error proofing bolts from defect
mother board and assemble them on new board.
8.
Assemble the new mother board and connect the cables
according to the label of the cable and the mark on the
board.
Small
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...