Vista
™
basic
Infusion Pump
Quick
Reference
Rx only
Vista™ and
Working Together for Excellence in Safety
are a trademarks of B. Braun Medical Inc.
©2002, B. Braun Medical Inc.
950853-Rev B 06/02
Alarm Troubleshooting
Air In Line!
• Check for air in air-in-line
detector.
• Check that tubing is properly
inserted in air-in-line detector.
• Remove air from line.
Battery
• Plug pump into wall outlet.
Too Low!
Door Open!
• Close door.
Dwnstream
• Open clamps.
Occlusion!
• Check for kinked tubing.
• Check IV site.
• Check occlusion pressure
setting.
Pump
• Pump is stopped.
Stopped!
• Either start infusion or press
to extend hold time
3 minutes.
Silence alarms by pressing
.
Refer to Instructions for Use for complete
operating instructions.
B. Braun Medical Inc.
824 Twelfth Avenue
Bethlehem, PA 18018
1-800-227-2862
www.bbraunusa.com
Program Mode
1.
Press OPTS
until
PROGRAM
MODE
is
displayed
2.
Press ON
.
3.
Enter the rate for Step 1 and press OK
.
4.
Enter the volume for Step 1 and press OK
.
5.
Repeat Steps 3 and 4 until all steps are
complete and press OK
.
6.
Review the Total VTBD and press OK
.
7.
Press START
.
Ramp Taper Mode
1.
Press OPTS
until
RAMP
TAPER
MODE
is
displayed
2.
Press ON
.
3.
Enter the total bag volume to be delivered and
press OK
.
4.
Enter the Ramp Up time in hours and minutes
and press OK
.
5.
Enter the Taper Down time in hours and minutes
and press OK
.
6.
Enter the Total Time for the infusion in hours
and minutes and press OK
.
7.
Review the calculated Max Rate and
press OK
.
8.
Press START
.