Troubleshooting
6-25
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0306514
Warning
Read MWL
server query
filter failed
Connectivity configuration
loading failure. Worklist query
filter setting missed. Please
restart the system. If this issue
persists, please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0306515
Warning
Load dose
summary icon
failed
Dose summary icon loading
failure. Display may be
abnormal. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0306516
Error
ImageDirectory
module register
OSGI service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0306517
Error
Call
DatabaseMana
gement service
failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0306518
Error
Call
ImageStorage
service failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0306519
Error
Call
CurrentExam
service failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0306520
Error
Call
MediaManage
ment service
failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0306521
Error
Subscribe event
failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0306522
Error
Bind name
service object
failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...