Brivo OEC 715/785/865/865 Mobile C-Arm X-Ray Product Service Manual
2-68
There are two connectors in GigE camera. One is industrial standard RJ-45 connector for GigE Vision
Protocol on Gigabit Ethernet, IP: 192.168.42.239. The other is HDB15 female connector which provides
power, RS422 and GPIO as following Table:
Pin
Name
Description
1
RSV
Reserved
2
LIH
Input, 12V logic Last image hold
3
Output, RS422 positive terminal of differential TXD signal
4
RXD485-
Input, RS22 negative terminal of differential RXD signal
5
GND
GND
6
RSV
Reserved
7
GND
GND
8
TXD485-
Output, RS422 negative terminal of differential TXD signal
9
Input, RS422 positive terminal of differential RXD signal
10
VCC
+12V supply voltage
11
GND
GND
12
RSV
Reserved
13
RSV
Reserved
14
FrameSync
Output, 12V logic Frame synchronization
15
GND
GND
Lens
GigE Camera AM1565 uses Computar M1614-MP lens. This lens has 16mm focal lens, manual iris F1.4-F16
and manual focus.
Computar M1614 Lens
GigE Camera AM1565 uses an interface flange for mounting itself into image intensifier. This interface flange
is independent part from lens. This interface flange provide three holes and these holes accommodate
errors of at least 2 mm for positioning the image center.
GigE Camera AM1565B use 3D 500025 lens. This lens has 16mm focal lens, manual iris F1.4-F16 and manual
focus.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...