background image

Pipette Accuracy Tester

PT-Series

 

AD-4212B-PT 
AD-4212A-PT 

FX-300

i

-PT

 

 

INSTRUCTION MANUAL 

1WMPD4001794 

Summary of Contents for PT Series

Page 1: ...Pipette Accuracy Tester PT Series AD 4212B PT AD 4212A PT FX 300i PT INSTRUCTION MANUAL 1WMPD4001794...

Page 2: ...by any means without the written permission of A D Company Limited The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement wit...

Page 3: ...ING A PC TO THE BALANCE 12 6 1 Connecting a PC 12 6 2 Checking COM Port Numbers 13 7 CALIBRATING THE BALANCE 14 7 1 AD 4212B PT AD 4212A PT 14 7 2 FX 300i PT 15 8 PREPARATION BEFORE USE 16 8 1 Filling...

Page 4: ...method for accuracy of the pipette volume For detailed information visit our web site 1 1 Unpacking the Tester NOTE The pipette accuracy tester is contained in a carrying case The case is only for car...

Page 5: ...manual for the tester Angle adjustment knob Capacity 5 mL 2 pcs Capacity 30 mL 2 pcs Sample cup holder 30 mL Evaporation trap base Sample cup holder 5 mL Shoulder belt Key Balance Fitting boss Small...

Page 6: ...at are previously entered the PC calculates the conversion factor Z factor from a mass value to a volumetric value and converts the mass value into a volumetric value automatically using the Z factor...

Page 7: ...at the humidity of the measurement environment be greater than 50 RH Water is poured into the groove of the evaporation trap This helps to maintain the humidity inside the evaporation trap high and re...

Page 8: ...3 PRECAUTIONS BEFORE USE Secure the display to the stand using the angle adjustment knobs Connect the display and the weighing unit using the connection cable 2 Install the evaporation trap base on th...

Page 9: ...the pipette insertion opening on the sloped surface is faced to allow easy operation 7 Install cap A on the evaporation trap Remove cap B from the sloped surface Lower part of the evaporation trap Low...

Page 10: ...rap base on the fitting boss 3 Install the sample cup holder 30 mL on the base When the 30 mL sample cup is used Insert the sample cup into the holder When the 5 mL sample cup is used Insert the sampl...

Page 11: ...that the pipette insertion opening on the sloped surface is faced to allow an easy operation 7 Install cap A on the evaporation trap Remove cap B from the sloped surface Lower part of the evaporation...

Page 12: ...e balance away from equipment which produces magnetic fields Install the balance in a stable place avoiding vibration and shock Corners of rooms close to pillars or walls far from a passage way are be...

Page 13: ...nator to remove the static electricity 5 INSTALLING WinCT Pipette For detailed information on the WinCT Pipette program refer to How to use WinCT Pipette pdf in the WinCT Pipette CD ROM For installing...

Page 14: ...Some PCs may require a driver installation even when the USB port used is changed The driver installation CD and the instruction manual are contained in the same package with the USB converter For in...

Page 15: ...anager 4 Double click Port COM LPT COM port numbers will be displayed When the USB converter is used ATEN USB to Serial Bridge COMx appears In the example below the COM port number is 4 The COM port f...

Page 16: ...Cal 0 is displayed press the PRINT key to enter the zero point 4 After the zero point is entered the calibration weight value 50 is displayed Remove the upper part of the evaporation trap while holdin...

Page 17: ...ht on the sample cup holder 30 mL Attach the evaporation trap on the base mating the tabs with the holes 5 Press the PRINT key to calibrate the weighing capacity When end is displayed remove the calib...

Page 18: ...When pouring water use much care not to spill water around the fitting boss of the balance weighing pan Attach the upper part of the evaporation trap to the lower part and place the evaporation trap...

Page 19: ...with the back up Using a coin rotate the battery cover anti clockwise to remove it 2 Remove the old battery 3 Install a new battery with the side flat side facing up 4 Using the coin reattach the batt...

Page 20: ...ve the sheath protector from the thermometer 2 Press the ON FAST button to turn the display on 3 Insert the sensor sheath into the water 4 The sampling interval is 10 seconds To speed this process pre...

Page 21: ...e values in memory before measurement Resetting the maximum and minimum temperatures stored in memory Resetting the maximum temperature Resetting the minimum temperature Returning to the current tempe...

Page 22: ...C to 100 C 3 0 C 50 C to 150 C 3 5 C 50 C to 200 C 4 0 C other Sampling interval Regular every 10 seconds FAST every 2 seconds Water protection JIS IPX7 Endures a 30 minute immersion in a water depth...

Page 23: ...imit of the adjustable volume range whichever is greater 2 50 of the maximum volume 3 100 of the maximum volume NOTE Pipette volumes and number of measurements can be changed 9 2 Judging the Results W...

Page 24: ...ssure even if changed in the range between 950 hPa and 1050 hPa influences measurements only by 0 01 Water temperature if changed by 5 C influences measurements by 0 11 To reduce errors caused by wate...

Page 25: ...or air conditioners or where no drafts exist Especially when using the AD 4212B PT AD 4212A PT with a minimum weighing value of 0 01 mg 0 1 mg use much care in choosing an installation site or cover t...

Page 26: ...for testing a liquid with a volume of less than 100 L Function table Functions related to pipette accuracy testing are as follows Class Item and Parameter Description 0 1 Cond Condition 2 Fast weighin...

Page 27: ...nces the measurement particularly when the volume of the test liquid is small For such a situation the display lock function is useful How this function works After the test liquid is dispensed from t...

Page 28: ...tion install the balance on the first floor in the corners of rooms close to pillars or walls far from a passage way Use the AD 1685 anti vibration table Use the optional AD 1683 DC static eliminator...

Page 29: ...ction manual WinCT Pipette CD ROM Carrying case with a shoulder belt and a key AC adapter Please confirm that the AC adapter type is correct for your local voltage and power receptacle type Power cons...

Page 30: ...000 0 8 16 0 3 6 0 5000 0 8 40 0 3 15 0 10000 0 6 60 0 3 30 0 0 1 AD 4212A PT Daily inspection simplified verification 1 6 FX 300i PT 4 The maximum volume selectable for adjustable volume pipettes 5 T...

Reviews: