Installation
3-43
Manual HLF
Select manual HLF mode. Take exposure and verify that a HLF exposure is taken. Verify that audible
tone beeps at normal rate.
Auto HLF and Digital Spot
This key is only effective in some countries. In the other countries, invalid key sound will be given
when pressing this key. High Level Fluoro and Digital Spot can be switched in Mode Setup
which is in system panel of software interface. High Level Fluoro Mode, Digital Spot Mode could be
enabled by pressing the right footswitch or handswitch.
The LED lights can reflect whether the Auto HLF or Digital Spot feature is available. Press this key
again to disable the High Level Fluoro Mode& Digital Spot Mode. Make sure the LED lights before
pressing the right footswitch or handswitch to make exposure.
Recursive Factor
Press recursive factor key
to select 2, 4, 8, 16 respectively to take an exposure. Verify the
amount of filtering can be increased and decreased. The noise in the image should be decreased as
the Recursive Factor is increased.
Auto/Manual Brightness/Contrast
Press the “Auto/Manual Brightness/Contrast” key to adjust the brightness and contrast level display.
The current contrast or brightness level is shown in a numeric value displayed on the bottom right
corner of the image.
Auto/Manual Smart Metal
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...