Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
7-38
2.
Remove the rear cover of the C-
Arm console.
3.
Loose the four screws and remove
the shield cover.
Do not touch the IGBT inverter module, unless the power of DC Bus capacitor has been
completely removed and they have been discharged completely.
4.
Disconnect all cables connected to IGBT
board and the cables to the fan.
5.
Remain the screws on the board after
disconnect the cable. Be careful not to
break the pins.
6.
Loose the screws that secured the IGBT
module and replace a new one.
7.
Connect all cables to IGBT module.
8.
Reassembly the shield cover and the rear
cover of console.
Make sure that all cables are connected before you power on the system. Otherwise,
the IGBT assembly will be damaged.
9.
Power on the system. Take exposure and verify the system works normally.
10.
Refer to section 4.2.1 to perform kVp accuracy check and perform KVp calibration if needed.
11.
Refer to section 4.3.1 to perform mA accuracy check and perform mA calibration if needed.
WARNING
CAUTION
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...