9-6
9.3.
Specification of Key Components
9.3.1.
X-Ray Generator
Nominal electric power
2.0kw (100kV, 20mA, 0.1s)
Maximum electric power
Normal Fluoro
440W (110kV, 4mA)
High Level Fluoro (HLF)
660W (110kV, 6mA; 55kV, 12mA)
Low dose
220W (110kV, 2mA)
Digital Spot
1760W (110kV, 16mA)
Radiography
2.2 kW (110kV, 20mA) @ 200V, 220V, 230V,
240V;
1.1 kW (110kV, 10mA) @ 100V, 110V, 120V.
High frequency generator
40kHz
Voltage
40kV~110kV
Normal Fluoro
When the X-ray tube current reaches its
highest value of 4mA, the obtainable highest
X-ray tube voltage is 110kV.
High Level Fluoro (HLF)
When the X-ray tube current reaches its
highest value of 12mA, the obtainable highest
X-ray tube voltage is 55kV.
Low dose
When the X-ray tube current reaches its
highest value of 2mA, the obtainable highest
X-ray tube voltage is 110kV.
Digital Spot
When the X-ray tube current reaches its
highest value of 16mA, the obtainable highest
X-ray tube voltage is 110kV.
Radiography:
20mA, stationary @ 200V, 220V, 230V, 240V;
the obtainable highest X-ray tube voltage is
110kV;
10mA, stationary @ 100V, 110V, 120V; the
obtainable highest X-ray tube voltage is 110kV.
Current
Normal Fluoro
0.1 mA~4 mA; when the X-ray tube voltage
reaches its highest value of 110kV, the
obtainable highest X-ray tube current is 4mA.
High Level Fluoro (HLF)
0.2 mA~12 mA; when the X-ray tube voltage
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...