Replacement
7-75
Emergency Switch (P/N: 5330849)
1.
Remove the rear and front cover of the C-Arm console.
2.
Pull out the spring clip and release the emergency switch.
3.
Disconnect the cable connected to mounting of the
emergency switch.
4.
Loosen the screws and remove the base and switch. Then
replace with a new one.
Ensure the switch head is buckled by the
support.
5.
Power on the system. Press the emergency switch, vertical
column operation will cease and X-rays will be disabled.
Rotate as arrow indication to release the emergency switch.
You can continue to use the system.
7.3.23.
Orbital Rolling Wheels Kit
(PN:
5599917
)
Spring clip
Buckle for switch head