Installation
3-63
Paste the multi language label on Removable
Grid Support.
Removable Grid Support
Carbon Protection Board
Grid for Brivo OEC 715, Brivo OEC 785
White dot is the direction mark for Grid.
The connected line of direction mark is
in45
°
with C-Arm, in Brivo OEC715/785.
Grid for Brivo OEC865
White dot is the direction mark for Grid.
The connected line of direction mark is 60
degree with C-Arm, in Brivo OEC 865.
Assembly and install removable grid
1.
Remove the Grid by releasing the 6 screws from image
intensifier.
2.
Install the Carbon Protection Board into image intensifier.
3.
Install Removable Grid Holder on image intensifier.
45
°
60
°