Calibration
4-3
4.2.
kV Measurement and Adjustment
4.2.1.
kV Measurement and Accuracy Check
The X-ray generator is composed of high voltage transformer, high voltage rectifier and X-ray tube in one
pressurized oil tank. kV value can be measured by kV meter. Position the kV meter in the center of the X-
ray beam 30 cm above the focal spot.
KV Accurancy requirement
Take exposure at kV values below in manual normal Fluoro mode. The error should be within
±
7%
ranges. Otherwise, perform the kV adjustment procedure in section 4.2.2 until all the kV values meet the
requirement below.
Exposure
mode
kVp
value
mA or mAs Set Value
Calculation
Error
Minimum
(kVp)
Maximum
(kVp)
Low Dose
99 kVp
2mA
7%
92.07
105.93
Normal
Fluoro
66kVp
4mA
7%
61.38
70.6
110 kVp
1mA
7%
102.3
117.7
High Level
Fluoro
55 KVp
4mA
7%
51.15
58.85
110 kVp
4mA
7%
102.3
117.7
Digital Spot
55 KVp
4mA
7%
51.15
57.75
80 kVp
4mA
7%
74.4
85.6
110 kVp
4mA
7%
102.3
117.7
Radiography 55 kVp
40mAs (For 100/110V/120V input)
80mAs (For
200V/220V/230V/240V input)
7%
51.15
58.85
80KVp
40mAs
7%
74.4
85.6
110 kVp
20mAs
7%
102.3
117.7
KV Accurancy requirement in Australia
Take exposure at kVp values below in normal fluoro mode. The error should be within ±5% ranges.
Otherwise, perform the kVp adjustment procedure in Section 4.2.2 until all the kVp values meet the
requirements below.
Exposure
mode
kVp
value
mA or mAs Set Value
Calculation
Error
Minimum
(kVp)
Maximum
(kVp)
Low Dose
99 kVp
2mA
5%
94.05
103.95
Normal
Fluoro
66kVp
4mA
5%
62.7
69.3
110 kVp
1mA
5%
104.5
115.5
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...