Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
7-42
7.3.10.
Parts in Computer and Peripheral
Perform the procedures below to replace the parts of computer.
1.
Power off the system and disconnect Workstation’s AC plug
from the AC receptacle.
2.
Remove the rear cover, side covers and front cover of the
Workstation.
3.
Unplug the cables from the computer. Loosen the screws of
the computer cover. Lift up and remove the cover.
4.
Replace the FRU parts (such as the hard disk, display card, Memory Bank, mother board, image
grabber card and Chassis Power) in the computer.
After the display card is replaced, if the software version is 011601 or higher version, it
is required to reinstall the workstation software (OS and Application Software) refer to
section 5.3.1
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...