Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-54
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0405508
Warning
SCU performs
storage
operation failed
DICOM storage failure. Please
make sure the network and
storage server configuration
are correct, and the server is
available. And try again. If this
issue persists, please call for
service.
Check DICM configuration.
0405509
Error
SCU sets locale
failed
System start failure. Set locale
failed. Please restart the
system. If this issue persists,
please call for service.
System unknown error
0405510
Error
SCU reports
ready failed
System start failure. Please
restart the system. If this issue
persists, please call for service.
make sure if omniNames
service failed, any module
crashed
0405701
Error
SCU module
resolve object
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service
make sure if omniNames
service failed, any module
crashed
0405702
Error
SCU module
Invoke corba
service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service
make sure if omniNames
service failed, any module
crashed
0405703
Error
SCU creates
thread failed
Execution thread exception.
Please restart the system. If
this issue persists, please call
for service.
System unknown error
0405704
Warning
SCU cancels
thread failed
Cancel execution thread
exception. Please restart the
system. If this issue persists,
please call for service.
System unknown error
0405705
Error
SCU allocates
memory failed
System run out of memory.
Please restart the system. If
this issue persists, please call
for service.
Check memory using state
0405706
Error
SCU module
register servant
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service
make sure if omniNames
service failed, any module
initialization failed
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...