Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
9-2
9.1.
Overview
The policy of GE Healthcare is one of continual product development and improvement.
For this reason, GE Healthcare reserves the right to change the operating characteristics and
specifications of newer products at any time, without prior notice, and without incurring any obligation
relating to previously manufactured items.
The specifications listed are limited to general performance and physical data. Specifications of optional
equipment provided by other manufacturers are given in the applicable manuals provided with those
options.
9.2.
System Specifications
9.2.1.
Safety Classification
Class I, Type B, IEC60601-1
Ordinary protection against ingress of water
Degree of Protection:
•
System: IPX0
•
Footswitch: IPX8
9.2.2.
Environmental Requirements
Operating Conditions:
Temperature: +10
°
C~+40
°
C (+50F~+104F)
Humidity: 20%~80% Relative, non-condensing
Atmospheric Pressure: 700hPa~1060hPa
Transportation and Storage:
Temperature: -20
°
C~+55
°
C (-4F~+131F)
Humidity: 10~95% Relative, non-condensing
Atmospheric Pressure: 500hPa~1060hPa
9.2.3.
Power Requirements
Voltage: AC 100V/110V/120V|200V|220V/230V/240V
Frequency: 50Hz/60Hz
Rated Momentary Line Current (shorter than 5s):
•
30A (100V/110V/120V)
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...