Technical Reference
9-7
reaches its highest value of 110kV, the
obtainable highest X-ray tube current is 6mA.
Low dose
0.1 mA~2 mA; when the X-ray tube voltage
reaches its highest value of 110kV, the
obtainable highest X-ray tube current is 2mA.
Digital Spot
0.2mA~16 mA; when the X-ray tube voltage
reaches its highest value of 110kV, the
obtainable highest X-ray tube current is 16mA.
Radiography
1mAs~80mAs @ 200V, 220V, 230V, 240V;
current: 20mA, stationary; the obtainable
highest X-ray tube voltage is 110kV;
1mAs~40mAs @ 100V, 110V, 120V; current:
10mA, stationary; the obtainable highest X-ray
tube voltage is 110kV.
Radiography
Voltage
40~110kVp
Value of mAs
1.0, 1.3, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8.0, 10, 13,
16, 20, 25, 32, 40, 50, 63, 80 mAs @ 200V, 220V,
230V, 240V;
1.0, 1.3, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8.0, 10, 13,
16, 20, 25, 32, 40 mAs @ 100V, 110V, 120V;
Pulse Fluoro
1/2/4/8pps
The lowest Current Time Product or the
combination of Loading Factors
40kVp, 1mAs
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...