Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-40
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0309508
Warning
Start setup
exception
Open Setup exception. Please
restart the system. If this issue
persists, please call for service.
system unknown error
0309509
Warning
Stop setup
exception
Close Setup exception. Please
restart the system. If this issue
persists, please call for service.
system unknown error
0309510
Warning
Read
configuration
file failed
Connectivity configuration
loading failure. Please restart
the system. If this issue
persists, please call for service.
check if configuration file
crashed
0309511
Warning
Init tab page
controller failed
Runtime exception. Please
restart the system. If this issue
persists, please call for service.
system unknown error
0309512
Warning
Read serial
number from
bios failed
Workstation hardware version
access exception. Hardware
information can’t be displayed.
Please restart the system. If
this issue persists, please call
for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0309513
Error
Register CORBA
service failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
make sure if omniNames
service failed, any module
initialization failed
0309514
Error
Unregister
CORBA service
failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
make sure if omniNames
service failed, check
omniOrb log file
0309515
Error
CORBA sync
notify failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
make sure if omniNames
service failed,
CorbaEventChannel
initialization failed
0309516
Error
Subscribe sync
event failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
make sure if omniNames
service failed,
CorbaEventChannel
initialization failed
0309517
Error
Unsubscribe
sync event
failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
make sure if omniNames
service failed,
CorbaEventChannel
initialization failed
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...