Replacement
7-65
Interconnect Cable Workstation
Emergency Switch
AC220V
3.
Remove the four screws securing the casting, and then remove the four screws securing the inner
socket to take out the head of GigE Ethernet cable.
4.
Replace the new cable and remove the four screws securing the inner socket to assemble the head of
GigE Ethernet cable.
5.
Fasten the four screws securing the inner socket and securing screws of casting.
6.
Connect all the cables and route as original.
Keep new cable routed as original to ensure the package of mainframe can be open and
closed smoothly.
7.
Recheck all connectors are correctly connected and cables are well routed.
8.
Power on the system, and take an exposure.
9.
If system works well, turn off system and assemble all covers.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...