Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
4-36
Definition for iris and Shutter position:
Make all iris and shutter leaves’ penumbra (One/both/all of the leaves) is just out of
image circular mask.
Methods for measuring Iris diameter:
Situation_1. In below left image:
Via opening Iris, two opposite leaves are just out of circular mask. Other leaves have
been out of image (Image intensifier output window). Measure the linear distance
between two leaves’ penumbra margins as the diameter of Iris.
Situation_2. In below right image,
Via opening Iris, only one leaves is just out of circular mask. Other leaves have been
out of light image (Image intensifier output window). So the linear distance between
two leaves’ penumbra can’t be measured. Herein measure the linear distance between
the leave margin and 180
°
light image (Image intensifier output window) margin as
the diameter of Iris.
Measurement margin
Circular mask
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...