Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
4-28
5.
Set to normal mode, fully open the collimator and make a shot.
6.
Align the BAFT tool center with image display center:
a.
Loosen the securing screws (4 screws for each horizontal and vertical adjustment).
b.
During a MAG 1 Fluoro exposure, adjust position adjustment screws (1 screw for each
horizontal and vertical adjustment) on the BAFT tool to make the BAFT center and
intersection of the two lines overlapped.
c.
Tighten the securing screws. Then make X-ray shot and double check the BAFT alignment.
Ensure BAFT tool is aligned perfectly before performance below calibration.
UBAT Installation
1.
Position the C-arm as Monoblock Up and Image Intensifier down.
2.
Position the UBAT (Universal Beam Alignment Template) on top of the Image Intensifier
as below.
8 Securing Screws
2 Alignment Screws
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...