Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0309518
Error
CORBA async
notify failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
make sure if omniNames
and notifd service failed,
CorbaEventChannel
initialization failed
0309519
Error
Subscribe
async event
failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
make sure if omniNames
and notifd service failed,
CorbaEventChannel
initialization failed
0309520
Error
Unsubscribe
async event
failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
make sure if omniNames
and notifd service failed,
CorbaEventChannel
initialization failed
0309521
Error
CORBA
communication
failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
make sure if omniNames
service failed, any module
crashed
0309522
Warning
Parse
configuration
file failed
Configuraiton loading
exception. Some configuration
may be missed. Please restart
the system. If this issue
persists, please call for service.
check if configuration file
(Admin/Exposure
contorl..)crashed
0309523
Warning
Export
configuration
file failed, no
enough space
or no
permission
Export configuration failure.
Please check if the USB disk is
writable and has enough
space, or change a disk.
system unknown error
0309524
Warning
Export log file
failed, no
enough space
or no
permission
Export log files failure. Please
check if the USB disk is writable
and has enough space, or
change a disk.
system unknown error
0309525
Warning
Create folder
failed, no
enough space
or no
permission
Export log files exception.
Please retry. If this issue
persists, please restart the
system.
IO error or disk error
0309526
Warning
Delete folder
failed
Operation exception. Please
retry. If this issue persists
please restart the system.
IO error or disk error
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...