background image

Operating Instructions

K4 Table top controller 4954

KaVo Elektrotechnisches Werk GmbH
Wangener Straße 78
D-88299 Leutkirch
Tel.: 0 75 61 / 86-0 • Fax: 0 75 61 / 86-23 3

1.000.3034 

WHD RB 

1

1

/00 GB  

01.33

Summary of Contents for K4

Page 1: ...Operating Instructions K4 Table top controller 4954 KaVo Elektrotechnisches Werk GmbH Wangener Stra e 78 D 88299 Leutkirch Tel 0 75 61 86 0 Fax 0 75 61 86 23 3 1 000 3034 WHD RB 11 00 GB 01 33...

Page 2: ......

Page 3: ...tential uses 3 A 1 5 Technical data 3 A 1 6 Combination Controller handpiece 3 A 2 Scope of delivery Accessories 4 A 3 Electrical connection 4 A 4 Installation and commissioning 5 A 5 Description 6 A...

Page 4: ...eed and maximum pressure of the tools according to tool manufacturer s instructions A 1 User information A 1 1 Meaning of the pictograms Situations where failure to follow the instructions may lead to...

Page 5: ...1000 30 000 min 1 In the event of extreme high frequen cy interfering fields speed variations up to max 20 may occur Rated voltages Voltage ranges 100 V 120 V 230 V 50 60 Hz Output voltage max 30 V D...

Page 6: ...ltage and rated frequency agree with the data on the rating plate and voltage selector switch If necessary make an adjustment Voltage adjustment By moving the voltage selector switch the country speci...

Page 7: ...mains plug into the easily accessible correctly installed earthed mains socket This should also be ensured when using an extension cable Insert plug of connecting cable of hand piece in unit socket an...

Page 8: ...top controller 4954 A 5 Description Malfunction display Speed preselection continuous Foot switch socket Unit socket 6 pin Voltage selector switch with voltage adjustment Mains switch 0 I Foot switch...

Page 9: ...4 see A 1 6 For correct operation of the handpiece used it is essential to follow the instruc tions for use 1 000 3902 supplied with the handpiece On the K4 controller 4954 press mains switch to posit...

Page 10: ...which the flash ing sequence begins again In the case of all faults the motor is switched off Some faults are reset when the foot switch is at stop serious errors can be reset only by switching off C...

Page 11: ...irect current chopper or measuring device U_wr I_wr faulty R Repair motor electronics C Inverter faulty automatic test on swit ching on R Repair motor electronics C Brake device faulty automatic test...

Page 12: ...HEET FOR USERS K4 TABLE TOP CONTROLLER 4954 10 K4 Table top controller 4954 1 000 1966 1 000 1967 0 200 6528 0 692 9021 0 200 6529 0 245 5015 0 223 2735 0 220 0469 1 000 0906 0 223 4142 Spare parts sh...

Page 13: ...necting instructions corrosion impurities in the air supply or chemical or electrical influences which are unusual or not admissible in accordance with KaVo s instructions The guarantee shall become n...

Reviews: