7. Messages
7-42
UniSite-xpi User Manual
Device Over-current Fault
Probable Cause
Solution
Improper device selected
Make sure the device selection matches the manufacturer and part
number of your device as precisely as possible. If it doesn’t, select the
proper characteristics and perform the operation again.
Note: Choosing a wrong manufacturer and/or part number (via the Select
device option from the Main Menu) causes the programmer to expect an
electronic ID different from your device’s ID.
Faulty device(s)
Load the suspect device (using the Load device option from the Main
Menu). If an overcurrent error occurs, load new devices. If other devices
load with no errors, a single device may be faulty. If no errors occur
during load operation, try to program another device (using the
Program Device option from the Main Menu). If other devices program
without error, a single device may be faulty.
If other devices also produce overcurrent errors during load or program
operations, check the date code. If failures occur only on devices with a
particular date code, while other parts are programmed or loaded
successfully, the problem is probably device related.
Note: If the problem appears to be device related, you may wish to notify the
device manufacturer.
Possible bug in software
associated with device tests
If other devices produce overcurrent error during load a program
operations, disable the Continuity check parameter displayed on the
Programming Parameters screen (via More/Configure
system/Edit/Programming options from the Main Menu).
If this error occurs during a programming operation but not during a
load operation, a device test may be causing the problem. Disable all
device checks (such as Device check and Illegal bit check) listed on the
Program Device screen by changing the appropriate fields from “y” to
“N.” Press the F4 key to display all parameters.
Improper algorithm applied by
programmer due to possible
bug in software
If this error occurs on devices with old and recent date codes, there may
be a software bug in the device’s programming algorithm.
Workaround:
On the programmer’s Device List, find an earlier version
of programmer software that supports your device, boot your
programmer with this version, and repeat the operation. If the operation
is successful, you’ve found a temporary workaround.
Note: Contact Data I/O Customer Support and report your findings.
Programmer hardware
problem
An overcurrent error occurring with different devices may indicate a
hardware problem. To determine whether a hardware malfunction
exists, perform a self-test (using the More/Self-test options from the
Main Menu) with no device sin the socket. If the self-test shows a
malfunction, make the necessary service arrangements.
Summary of Contents for UniSite-xpi
Page 2: ...981 0014 UniSite Programming System User Manual Also Covering Legacy UniSite Programmers ...
Page 72: ...2 Setup and Installation 2 42 UniSite xpi User Manual ...
Page 134: ...3 Getting Started 3 62 UniSite xpi User Manual ...
Page 238: ...5 Computer Remote Control 5 10 UniSite xpi User Manual ...
Page 296: ...6 Translation Formats 6 58 UniSite xpi User Manual ...
Page 402: ...Keep Current Subscription Service 6 Keep Current ...