5. Computer Remote Control
UniSite-xpi User Manual
5-3
If you wish to have the programmer power up in CRC, perform the
following steps:
1.
Press F1 to get to the Main Menu.
2.
Type M to select the More Commands menu.
3.
Press C to select the Configure System menu.
4.
Press E to select Edit from the Configure Systems menu.
5.
Press I to select Interface from the Edit menu. The programmer
displays the interface parameters.
6.
Move the cursor to the Power on CRC field and press Y. CRC is now
selected. The following steps in this procedure save CRC mode as a
powerup system parameter.
7.
Press F2 two times to return to the Configure System Parameters
menu.
8.
Press S to select Save from the Configure System Parameters menu.
The screen displays the Save System Parameters menu.
9.
Type 1 and press E
NTER
to select the Powerup Defaults file as the one
where system parameters will be saved.
10. Press E
NTER
again so that the selection will be saved to the disk. The
next time you power up the programmer, it will enter CRC mode
automatically.
Which Interface Mode?
You can operate the programmer in one of two interface modes: Terminal
and CRC. In Terminal mode you use screens and menus to interact with
the programmer. In CRC mode you send single-line commands to the
programmer, and the programmer responds with single line prompts,
responses, and error codes.
Note: Terminal mode operations may be run from either the Terminal port or the
Remote port. CRC mode operations must be run from the Remote port;
CRC will not work on the Terminal port.
Depending on the equipment you have connected to the programmer,
and on the settings of the User Menu Port and Power-on CRC
parameters, you can select which mode is available on which port. (The
User Menu Port parameter is found on the More Commands/Configure
System/Edit/Communication Parameters screen.)
Factory defaults for the programmer are Terminal mode commands sent
through the Terminal port and CRC mode commands sent through the
Remote port. The factory default for power-up state is Terminal mode.
Exiting CRC Mode
Press S
HIFT
+ Z + E
NTER
to exit CRC from an ASCII terminal on the
Terminal port.
From a remote computer, send the Z, E
NTER
command. If you exit remote
mode using the Z, E
NTER
command, the programmer’s parameters are set
to what they were before you entered remote mode.
Summary of Contents for UniSite-xpi
Page 2: ...981 0014 UniSite Programming System User Manual Also Covering Legacy UniSite Programmers ...
Page 72: ...2 Setup and Installation 2 42 UniSite xpi User Manual ...
Page 134: ...3 Getting Started 3 62 UniSite xpi User Manual ...
Page 238: ...5 Computer Remote Control 5 10 UniSite xpi User Manual ...
Page 296: ...6 Translation Formats 6 58 UniSite xpi User Manual ...
Page 402: ...Keep Current Subscription Service 6 Keep Current ...