7. Messages
7-34
UniSite-xpi User Manual
TEST ABORTED: Both calibration & PCU must pass before Pin driver test.
If you selected self-tests for the Pin Driver boards, and either the calibration or the Pin Control Unit did
not pass the self-test, this message will appear. Perform separate self-tests on both the calibration and the
Pin Control Unit prior to testing the Pin Driver boards.
TEST ABORTED: Both waveform board & PCU must pass before Pin driver test.
If you selected self-tests for the Pin Driver boards, and either the Waveform board or the Pin Control Unit
did not pass the self-test, this message will appear. Perform separate self-tests on both the Waveform
board and the Pin Control Unit prior to testing the Pin Driver boards.
TEST ABORTED: Calibration test must pass before PCU can be tested.
This message appears when you have attempted to perform a self-test on the Pin Control Unit (PCU)
before the Calibration test has passed. Perform the Calibration test prior to testing the PCU.
TEST ABORTED: Diskette not installed.
If you selected either drive A or drive B from the self-test screen, but a formatted disk is not in that drive,
this message will appear. Insert a formatted disk into the drive and try the test again.
TEST ABORTED: Drive B not installed.
If you selected drive B from the self-test screen, but drive B is not installed, this message will appear.
Install the disk drive or select drive A.
TEST ABORTED: FSM and PSM not installed.
If you selected a self-test for the Waveform board, Pin Control Unit, or Pin Driver board, and the PSM or
FSM is not installed, this message will appear. Install on or both of the modules.
TEST ABORTED: FSM not installed.
If you selected the FSM test from the self-test screen, and an FSM is not installed, this message will appear.
Install the module and try the self-test again.
TEST ABORTED: PSM not installed.
If you selected the PSM test from the self-test screen and a PSM is not installed, this message will appear.
Install the module and try the self-test again.
TEST ABORTED: Selected pin driver board is not installed.
The Pin Driver board you attempted to test is not installed. Install the board or select a different board to
test.
TEST ABORTED: Socket must be empty to perform test.
The device socket(s) must be empty in order to perform the self-test. Remove any device(s) which may be
present in the device socket(s). If this message is being reported while the sockets are empty, the
programmer hardware may be defective. Contact your local Data I/O Service Center if the problem
persists.
TEST ABORTED: Unknown adapter installed.
This message appears if the adapter installed is not recognized by the programmer. Update your
system/algorithm disk.
Summary of Contents for UniSite-xpi
Page 2: ...981 0014 UniSite Programming System User Manual Also Covering Legacy UniSite Programmers ...
Page 72: ...2 Setup and Installation 2 42 UniSite xpi User Manual ...
Page 134: ...3 Getting Started 3 62 UniSite xpi User Manual ...
Page 238: ...5 Computer Remote Control 5 10 UniSite xpi User Manual ...
Page 296: ...6 Translation Formats 6 58 UniSite xpi User Manual ...
Page 402: ...Keep Current Subscription Service 6 Keep Current ...