7-(7)
Troubleshooting
Section 7-2
Failed to load data.
Mount the Memory Card properly and re-execute.
Cannot delete directory having files.
Delete the files in the directory to be deleted or move the files to
another directory.
Reason: Only empty directories can be deleted.
Refer to
'HOHWLQJ)LOHVDQG'LUHFWRULHV
.
Directory name is not specified.
Specify the directory name and press the ENT Key.
Directory name contains unusable
character.
Check the directory name. The following characters cannot be used:
¥
/ , : ; * ?
< > ! & . SPC (space)
Same directory name has already
existed.
Specify a directory name that is not being used.
Same file name has already existed.
Specify a file name that is not being used in the directory.
Calculated magnification highers. Cor-
rect magnification range: 0.001 to
9.999.
Check for the following items, set the magnification again, and re-exe-
cute calibration:
- Widen the point or sampling width.
- Shorten the length of the actual coordinates.
Reason: The Camera magnification is not between 0.00100 and
9.99999.
Refer to
,QSXWWLQJ&DPHUD,PDJHV
Calculated magnification lowers. Cor-
rect magnification range: 0.001 to
9.999.
Password is incorrect.
Enter the correct password.
Refer to
6HWWLQJ3DVVZRUGV
and
1RUPDO6HULDO,QWHUIDFH
.
No Backup Data in memory card.
Mount the Memory Card on which the backup data was saved to Mem-
ory Card slot 1 (C1).
Refer to
/RDGLQJ'DWDRQSDJH
.
The backup data is not correct.
This backup data is damaged and cannot be loaded.
Date/time settings error.
Set the correct date and time.
Reason: A date not in the calendar or an invalid time has been input.
Refer to
6HWWLQJWKH&DOHQGDU'DWHDQG7LPH'DWH7LPH
.
File name is not specified.
Specify the file name and then press the ENT Key.
File name contains unusable charac-
ters.
Check the directory name. The following characters cannot be used:
¥
/ , : ; * ?
< > ! & . SPC (space)
Fan error.*
Turn OFF the Controller power supply and check that nothing is
obstructing the fan operation.
If the error message continues to be displayed when the Controller is
restarted, the Controller itself may be malfunctioning. Contact your
OMRON representative.
Cannot select image 1 when Defect
(Deviation) is registered.
To use Image 1 for measurement, delete the region to which density
deviation is set.
Refer to
'HQVLW\'HIHFWV
.
Contents of memory card is incorrect.
Cannot execute Select items and
Install.
Turn OFF the Controller power supply, re-install the F250-UME, and
restart the Controller. If the error remains even after the Controller is
started again, the F250-UME may be malfunctioning. Contact your
OMRON representative.
Memory card access error.
Check for the following items:
- The recommended Memory Card has not been inserted.
- Use another Memory Card.
- The Memory Card has not been formatted.
Refer to the Setup Manual.
Memory card is not inserted.
Insert the recommended Memory Card.
Refer to the Setup Manual.
Insufficient free space of memory card
- Delete unnecessary files to create free space.
- Replace the Memory Card with a bigger capacity card.
Refer to
0HPRU\&DUG2SHUDWLRQV
.
Error message
Causes and remedies
Содержание F250-UME
Страница 2: ...80 SSOLFDWLRQ 6RIWZDUH 2SHUDWLRQ 0DQXDO 5HYLVHG HFHPEHU...
Страница 3: ...iii...
Страница 337: ...2 15 4 EC Defect Section 2 15 4 Press the ENT Key The settings will be registered and the screen in 1 will return...
Страница 531: ...2 23 4 ECM Search Section 2 23 4 Press the ENT Key The settings will be registered and the screen in 1 will return...
Страница 748: ...2 49 5 Display Line Display Box Display Circle and Display Cursor Section 2 49...
Страница 798: ...4 44 Memory Card Operations Section 4 6...
Страница 998: ...7 18 Menu Hierarchy Section 7 6...