4-(6)
Changing Scenes and Scene Groups
Section 4-1
4-1-3
Initializing Measurement Conditions: Clearing Scenes
Use the following procedure to clear measurement conditions set in Set Mode,
and to return scenes to their default settings. This section explains the proce-
dure performed separately for each scene.
SeeAlso
Refer to
8VLQJ 6FHQH *URXS )XQFWLRQ
for information on initializing
whole scene groups.
1.
Move the cursor to Scn and press the ENT Key.
2.
Move the cursor to the number of the scene to be cleared.
3.
Press the SHIFT+ESC Keys.
The Copy/Clear/Comment Menu will be displayed.
4.
Select Clear.
A confirmation message will be displayed.
5.
Select Execute.
The scene will be cleared.
S+ESC:Edit S+ENT:Scn Group
Image 0 freeze
---- ---ms
0.Scn 0
1.Scn 1
2.Scn 2
3.Scn 3
4.Scn 4
5.Scn 5
6.Scn 6
7.Scn 7
↑
↓
0.Scn 0
MON
Copy
Clear
Comment
This scene data
will be cleared
Execute
Cancel