2-42-(2)
Normal Data
Section 2-42
2-42-1
Selecting Output Formats
Select the format for outputting data to a serial interface.
ASCII Format
The following items can be set.
Note The asterisk (*) indicates the default setting.
Binary Format
The measurement result multiplied by 1,000 is output as four consecutive
bytes as one unit of data.
Item
Details
Output form
Select ASCII*.
Digits of integer
Set the number of digits for integer output. When 0 is selected, all
digits of data are output without leading zeros. Select a value
between 0 and 8*.
If the data has more digits than the number set, 9 will be output
for each digit.
Example: If 4 digits have been set and the data is “15619,” the
output will be “9999.”
Digits of decimal
Set the number of digits for decimal output. When 0 is selected,
the decimal places are rounded off. Select a value between 0 and
3*.
Minus
Select what sign will be displayed for negative values. Select
either
-
* or 8.
Field separator
Select the separator between data fields.
Select none, comma*, tab, space, or delimiter.
Record separator
Select the separator between data records.
Select none, comma, tab, space, or delimiter*.
0 Suppress
Select how to adjust any empty digits to the left of output data.
Select either ON or OFF*:
ON: Inserts a 0 into empty digits.
OFF: Inserts a space into empty digits.
Example: If “5” has been set for Digits of integer and “3” for Digits
of decimal, and the data is 100.000, the following output will be
made.
When ON is set: 00100.000
When OFF is set: (space)100.000 (i.e., a space is inserted).
Item
Details
Output form
Select Binary.
Digits of integer
These settings are only for ASCII format. No settings are required
for binary format.
Digits of decimal
Minus
Field separator
Record separator
0 Suppress
Содержание F250-UME
Страница 2: ...80 SSOLFDWLRQ 6RIWZDUH 2SHUDWLRQ 0DQXDO 5HYLVHG HFHPEHU...
Страница 3: ...iii...
Страница 337: ...2 15 4 EC Defect Section 2 15 4 Press the ENT Key The settings will be registered and the screen in 1 will return...
Страница 531: ...2 23 4 ECM Search Section 2 23 4 Press the ENT Key The settings will be registered and the screen in 1 will return...
Страница 748: ...2 49 5 Display Line Display Box Display Circle and Display Cursor Section 2 49...
Страница 798: ...4 44 Memory Card Operations Section 4 6...
Страница 998: ...7 18 Menu Hierarchy Section 7 6...