458
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
21
:MEASure Commands
:MEASure:VRMS
[<interval>]
[,<type>][,<source>]
(see
:MEASure:VRMS?
[<interval>]
[,<type>][,<source>]
<interval> ::= {CYCLe | DISPlay}
<type> ::= {AC | DC}
<source> ::= {CHANnel<n> |
FUNCtion<m> | MATH<m> |
WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<m> ::= 1 to (# math functions)
in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<return_value> ::= calculated dc
RMS voltage in NR3 format
:MEASure:VTOP
[<source>] (see
:MEASure:VTOP?
[<source>] (see
<source> ::= {CHANnel<n> |
FUNCtion<m> | MATH<m> |
WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<m> ::= 1 to (# math functions)
in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<return_value> ::= voltage at the
top of the waveform in NR3 format
:MEASure:WINDow
<type> (see
:MEASure:WINDow? (see
<type> ::= {MAIN | ZOOM | AUTO |
GATE}
:MEASure:XMAX
[<source>] (see
:MEASure:XMAX?
[<source>] (see
<source> ::= {CHANnel<n> |
FUNCtion<m> | FFT | MATH<m> |
WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<m> ::= 1 to (# math functions)
in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<return_value> ::= horizontal
value of the maximum in NR3
format
Table 96
:MEASure Commands Summary (continued)
Command
Query
Options and Query Returns