:WGEN<w> Commands
33
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
1227
:WGEN<w>:MODulation:AM:DEPTh
Command Syntax
:WGEN<w>:MODulation:AM:DEPTh <percent>
<w> ::= 1 in NR1 format
<percent> ::= AM depth percentage from 0% to 100% in NR1 format
The :WGEN<w>:MODulation:AM:DEPTh command specifies the amount of
amplitude modulation.
AM Depth refers to the portion of the amplitude range that will be used by the
modulation. For example, a depth setting of 80% causes the output amplitude to
vary from 10% to 90% (90% – 10% = 80%) of the original amplitude as the
modulating signal goes from its minimum to maximum amplitude.
Query Syntax
:WGEN<w>:MODulation:AM:DEPTh?
The :WGEN<w>:MODulation:AM:DEPTh? query returns the AM depth percentage
setting.
Return Format
<percent><NL>
<percent> ::= AM depth percentage from 0% to 100% in NR1 format
See Also
•
":WGEN<w>:MODulation:AM:FREQuency"
•
":WGEN<w>:MODulation:FM:DEViation"
•
":WGEN<w>:MODulation:FM:FREQuency"
•
":WGEN<w>:MODulation:FSKey:FREQuency"
•
":WGEN<w>:MODulation:FSKey:RATE"
•
":WGEN<w>:MODulation:FUNCtion"
•
":WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry"
•
•