1102
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
31
:TRIGger Commands
:TRIGger:GLITch:QUALifier
Command Syntax
:TRIGger:GLITch:QUALifier <operator>
<operator> ::= {GREaterthan | LESSthan | RANGe}
This command sets the mode of operation of the glitch pulse width trigger. The
oscilloscope can trigger on a pulse width that is greater than a time value, less
than a time value, or within a range of time values.
Query Syntax
:TRIGger:GLITch:QUALifier?
The :TRIGger:GLITch:QUALifier? query returns the glitch pulse width qualifier.
Return Format
<operator><NL>
<operator> ::= {GRE | LESS | RANG}
See Also
•
"Introduction to :TRIGger Commands"
•
•