Commands Quick Reference
5
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
165
Table 63
:TRIGger:ZONE Commands Summary
Command
Query
Options and Query Returns
:TRIGger:ZONE:SOURce
<source> (see
)
:TRIGger:ZONE:SOURce?
<source> ::= {CHANnel<n>}
<n> ::= 1 to (# analog channels)
in NR1 format
:TRIGger:ZONE:STATe
{{0 | OFF} | {1 |
ON}} (see
)
:TRIGger:ZONE:STATe?
{0 | 1}
:TRIGger:ZONE<n>:MODE
<mode> (see
)
:TRIGger:ZONE<n>:MODE
<mode> ::= {INTersect |
NOTintersect}
<n> ::= 1-2 in NR1 format
:TRIGger:ZONE<n>:PLAC
ement <width>,
<height>, <x_center>,
<y_center> (see
)
:TRIGger:ZONE<n>:PLAC
<width> ::= width of zone in
seconds
<height> ::= height of zone in
volts
<x_center> ::= center of zone in
seconds
<y_center> ::= center of zone in
volts
<n> ::= 1-2 in NR1 format
n/a
:TRIGger:ZONE<n>:VALi
<value> ::= {VALid | INValid |
OSCReen}
<n> ::= 1-2 in NR1 format
:TRIGger:ZONE<n>:STAT
e {{0 | OFF} | {1 |
ON}} (see
)
:TRIGger:ZONE<n>:STAT
{0 | 1}
<n> ::= 1-2 in NR1 format
Table 64
:WAVeform Commands Summary
Command
Query
Options and Query Returns
:WAVeform:BYTeorder
<value> (see
)
:WAVeform:BYTeorder?
<value> ::= {LSBFirst | MSBFirst}
n/a
:WAVeform:COUNt? (see
)
<count> ::= an integer from 1 to
65536 in NR1 format