:TIMebase:SCALe
Command Syntax
:TIMebase:SCALe <scale_value>
<scale_value> ::= time/div in seconds in NR3 format
The :TIMebase:SCALe command sets the horizontal scale or units per division for
the main window.
Query Syntax
:TIMebase:SCALe?
The :TIMebase:SCALe? query returns the current horizontal scale setting in
seconds per division for the main window.
Return Format
<scale_value><NL>
<scale_value> ::= time/div in seconds in NR3 format