1130
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
31
:TRIGger Commands
:TRIGger:PXI:MSLot
Command Syntax
:TRIGger:PXI:MSLot <slot_number>
<slot_number> ::= 2 to (# chassis slots) in NR1 format
When this PXIe oscilloscope is a slave, the :TRIGger:PXI:MSLot command specifies
the chassis slot number of the master module.
It is important to identify the master slot number on each slave so that the proper
set of calibration factors can be used when aligning waveforms.
Calibration factors are created and saved using the KtScopeCal.exe application on
the chassis controller.
Query Syntax
:TRIGger:PXI:MSLot?
The :TRIGger:PXI:MSLot? query returns the master slot number setting.
Return Format
<slot_number><NL>
<slot_number> ::= 2 to (# chassis slots) in NR1 format
See Also
•
":TRIGger:PXI:MALine<n>:ENABle"
•
•
•
•
•