758
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
27
:SBUS<n> Commands
:SBUS<n>:CAN:SOURce
<source> (see
:SBUS<n>:CAN:SOURce?
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels)
in NR1 format
:SBUS<n>:CAN:TRIGger
<condition> (see
:SBUS<n>:CAN:TRIGger?
<condition> ::= {SOF | EOF |
IDData | DATA | FDData | IDRemote
| IDEither | ERRor | ACKerror |
FORMerror | STUFferror | CRCerror
| SPECerror | ALLerrors | BRSBit
| CRCDbit | EBActive | EBPassive
| OVERload | MESSage | MSIGnal |
FDMSignal}
:SBUS<n>:CAN:TRIGger:
IDFilter {{0 | OFF} |
{1 | ON}} (see
:SBUS<n>:CAN:TRIGger:
IDFilter? (see
{0 | 1}
:SBUS<n>:CAN:TRIGger:
PATTern:DATA <string>
(see
:SBUS<n>:CAN:TRIGger:
PATTern:DATA? (see
<string> ::= "nn...n" where n ::=
{0 | 1 | X | $}
<string ::= "0xnn...n" where n
::= {0,..,9 | A,..,F | X | $}
:SBUS<n>:CAN:TRIGger:
PATTern:DATA:DLC
<dlc> (see
:SBUS<n>:CAN:TRIGger:
PATTern:DATA:DLC?
<dlc> ::= integer between -1
(don't care) and 64, in NR1
format.
:SBUS<n>:CAN:TRIGger:
PATTern:DATA:LENGth
<length> (see
:SBUS<n>:CAN:TRIGger:
PATTern:DATA:LENGth?
<length> ::= integer from 1 to 8
in NR1 format
:SBUS<n>:CAN:TRIGger:
PATTern:DATA:STARt
<start> (see
:SBUS<n>:CAN:TRIGger:
PATTern:DATA:STARt?
<start> ::= integer between 0 and
63, in NR1 format.
:SBUS<n>:CAN:TRIGger:
PATTern:ID <string>
(see
:SBUS<n>:CAN:TRIGger:
PATTern:ID? (see
<string> ::= "nn...n" where n ::=
{0 | 1 | X | $}
<string ::= "0xnn...n" where n
::= {0,..,9 | A,..,F | X | $}
:SBUS<n>:CAN:TRIGger:
PATTern:ID:MODE
<value> (see
:SBUS<n>:CAN:TRIGger:
PATTern:ID:MODE? (see
<value> ::= {STANdard | EXTended}
:SBUS<n>:CAN:TRIGger:
SYMBolic:MESSage
<name> (see
:SBUS<n>:CAN:TRIGger:
SYMBolic:MESSage?
<name> ::= quoted ASCII string
Table 104
:SBUS<n>:CAN Commands Summary (continued)
Command
Query
Options and Query Returns