1054
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
30
:TIMebase Commands
:TIMebase:REFClock
Command Syntax
:TIMebase:REFClock <value>
<value> ::= {{1 | ON} | {0 | OFF}
The :TIMebase:REFClock command enables or disables the REF I/O MMCX
connector located on the front panel of the oscilloscope.
The REF I/O MMCX connector can be used as an input for the oscilloscope's
reference clock (instead of the internal 10 MHz reference), or it can be used to
output the internal 10 MHz reference clock when synchronizing multiple
instruments (see
The :TIMebase:REFClock ON command enables the REF I/O MMCX and sets the
reference signal mode to IN. The :TIMebase:REFClock OFF command disables the
REF I/O MMCX (the same as setting the reference signal mode to OFF).
Query Syntax
:TIMebase:REFClock?
The :TIMebase:REFClock? query returns the current state of the 10 MHz reference
signal mode. A "1" indicates that the REF I/O input is enabled (on), and a "0"
indicates that either the REF I/O MMCX is disabled (off) or that it is set as an
output (by the :ACQuire:RSIGnal command).
Return Format
<value><NL>
<value> ::= {0 | 1}
See Also
•