Commands Quick Reference
5
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
167
n/a
:WAVeform:PREamble?
<preamble_block> ::= <format
NR1>, <type NR1>,<points
NR1>,<count NR1>, <xincrement
NR3>, <xorigin NR3>, <xreference
NR1>,<yincrement NR3>, <yorigin
NR3>, <yreference NR1>
<format> ::= an integer in NR1
format:
• 0 for BYTE format
• 1 for WORD format
• 2 for ASCii format
<type> ::= an integer in NR1
format:
• 0 for NORMal type
• 1 for PEAK detect type
• 3 for AVERage type
• 4 for HRESolution type
<count> ::= Average count, or 1
if PEAK detect type or NORMal; an
integer in NR1 format
:WAVeform:SEGMented:A
LL {{0 | OFF} | {1 |
ON}} (see
)
:WAVeform:SEGMented:A
<setting> ::= {0 | 1}
n/a
:WAVeform:SEGMented:C
<count> ::= an integer from 2 to
1000 in NR1 format
n/a
:WAVeform:SEGMented:T
<time_tag> ::= in NR3 format
n/a
:WAVeform:SEGMented:X
LISt? <xlist_type>
<xlist_type> ::= {RELXorigin |
ABSXorigin | TTAG}
<return_value> ::= X-info for all
segments
:WAVeform:SOURce
<source> (see
)
:WAVeform:SOURce?
<source> ::= {CHANnel<n> |
FUNCtion<m> | MATH<m> | FFT |
WMEMory<r> | SBUS{1 | 2}}
<n> ::= 1 to (# analog channels)
in NR1 format
<m> ::= 1 to (# math functions)
in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
Table 64
:WAVeform Commands Summary (continued)
Command
Query
Options and Query Returns