696
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
25
:RECall Commands
:RECall:DBC[:STARt]
Command Syntax
:RECall:DBC[:STARt] [<file_name>] [, <serialbus>]
<file_name> ::= quoted ASCII string
<serialbus> ::= {SBUS<n>}
<n> ::= 1 to (# of serial bus) in NR1 format
The :RECall:DBC[:STARt] command loads a CAN DBC (communication database)
symbolic data file into the oscilloscope.
The <serialbus> parameter specifies which serial decode waveform the CAN
symbolic data will be loaded for.
See Also
•
"Introduction to :RECall Commands"
•
•
•
":SBUS<n>:CAN:TRIGger:SYMBolic:MESSage"
•
":SBUS<n>:CAN:TRIGger:SYMBolic:SIGNal"
•
":SBUS<n>:CAN:TRIGger:SYMBolic:VALue"
•
•
":SEARch:SERial:CAN:SYMBolic:MESSage"
•
":SEARch:SERial:CAN:SYMBolic:SIGNal"
•
":SEARch:SERial:CAN:SYMBolic:VALue"
NOTE
If a file extension is provided as part of a specified <file_name>, it must be ".dbc".