:SBUS<n>:SENT:SIGNal<s>:MULTiplier
Command Syntax
:SBUS<n>:SENT:SIGNal<s>:MULTiplier <multiplier>
<s> ::= 1-6, in NR1 format.
<multiplier> ::= from 1-24, in NR3 format.
When the display mode setting is SYMBolic (see :SBUS<n>:SENT:DISPlay), the
:SBUS<n>:SENT:SIGNal<s>:MULTiplier command specifies the multiplier to be
used in calculating a physical value displayed for a Fast Channel Signal.
• PhysicalValue = (Multiplier * SignalValueAsUnsignedInteger) + Offset
Query Syntax
:SBUS<n>:SENT:SIGNal<s>:MULTiplier?
The :SBUS<n>:SENT:SIGNal<s>:MULTiplier? query returns the multiplier value for
the Fast Channel Signal.
Return Format
<multiplier><NL>
<multiplier> ::= from 1-24, in NR3 format.