:TRIGger Commands
:TRIGger:PXI:MALine<n>:ENABle
Command Syntax
:TRIGger:PXI:MALine<n>:ENABle <setting>
<setting> ::= {{0 | OFF} | {1 | ON}}
<n> ::= 0 to (# chassis lines - 1) in NR1 format
When this PXIe oscilloscope is the master, the :TRIGger:PXI:MALine<n>:ENABle
command enables the chassis PXI_TRIG lines (0-7) on which to wait for "armed"
signals before triggering.
Each slave module must use a different PXI_TRIG line for its arm signal.
The master module will wait until all slave modules are armed before triggering.
If the modules are in different chassis trigger bus segments (which are isolated by
default), you must use the chassis Soft Front Panel (SFP) application to connect
the selected PXI_TRIG lines between trigger bus segments.
Query Syntax
:TRIGger:PXI:MALine<n>:ENABle?
The :TRIGger:PXI:MALine<n>:ENABle? query returns whether the specified chassis
arm line is enabled.
Return Format
<setting><NL>
<setting> ::= {0 | 1}