1318
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
37
Status Reporting
The status register bits are described in more detail in the following tables:
•
•
•
•
•
•
•
•
Run bit set if oscilloscope not stopped
Wait
Trig
Run
Arm
Reg
AER ?
:OPERegister:CONDition?
Operation Status Condition Register
:OPEE
:OPEE?
Operation Status Enable (Mask) Register
1
0
2
3
4
5
6
7
OPER
TRG
M AV
ESB
RQS/
MSS
TRG
Reg
TER?
Trigger Event Register
*SRE
*SRE?
Service Request Enable (Mask) Register
*STB?
Status Byte Register
Ou tp u t
Queue
SR Q
*ESR?
(Standard) Event Status Register
*ESE
*ESE?
(Standard) Event Status Enable (Mask) Register
PON
URQ
EXE
DDE
QYE
RQL
OPC
CM E
OVLR
USR
MSG
OR
1
0
2
3
4
5
6
7
OR
1
0
2
3
4
5
6
7
9
8
10
11
12
13
14
15
Se rvic e Re qu est
OR
Wait
Trig
Run
:OPERegister[:EVENt]?
Operation Status Event Register
OVLR
11
5
3
From Overload
Event Registers
From Mask Test
Event Registers
MTE
MTE
9
IOF
IOC
IOF
IOC
13
14
From Hardware
Event Registers
12
HWE
HWE
RUI
Enab
RUI
Enab
Set if remote user interface is enabled