Root (:) Commands
7
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
237
:OVLRegister (Overload Event Register)
Query Syntax
:OVLRegister?
The :OVLRegister query returns the overload protection value stored in the
Overload Event Register (OVLR). If an overvoltage is sensed on a 50
Ω
input, the
input will automatically switch to 1 M
Ω
input impedance. A "1" indicates an
overload has occurred.
NOTE
You can set analog channel input impedance to 50
Ω
on the 300 MHz, 500 MHz, and 1 GHz
bandwidth oscilloscope models. On these same bandwidth models, if there are only two
analog channels, you can also set external trigger input impedance to 50
Ω
.
Table 82
Overload Event Register (OVLR)
Bit
Description
When Set (1 = High = True), Indicates:
15-10
---
(Not used.)
9
Channel 4 Fault
Fault has occurred on Channel 4 input.
8
Channel 3 Fault
Fault has occurred on Channel 3 input.
7
Channel 2 Fault
Fault has occurred on Channel 2 input.
6
Channel 1 Fault
Fault has occurred on Channel 1 input.
5-4
---
(Not used.)
3
Channel 4 OVL
Overload has occurred on Channel 4 input.
2
Channel 3 OVL
Overload has occurred on Channel 3 input.
1
Channel 2 OVL
Overload has occurred on Channel 2 input.
0
Channel 1 OVL
Overload has occurred on Channel 1 input.
9
8
10
11
12
13
14
15
To OVLR bit in
Operati on Status
Register
OR
1
0
2
3
4
5
6
7
:OVLR?
Overload Event Register
:OVL
:OVL?
Overload Event Enable
(Mask) Register
Chan2
OVL
Chan1
OVL
Chan3
OVL
Chan4
OVL
Chan2
Fault
Chan1
Fault
Chan3
Fault
Chan4
Fault