:TIMebase Commands
:TIMebase:POSition
Command Syntax
:TIMebase:POSition <pos>
<pos> ::= time in seconds from the trigger to the display reference
in NR3 format
The :TIMebase:POSition command sets the time interval between the trigger event
and the display reference point on the screen. The display reference point is either
left, right, or center and is set with the :TIMebase:REFerence command. The
maximum position value depends on the time/division settings.
Query Syntax
:TIMebase:POSition?
The :TIMebase:POSition? query returns the current time from the trigger to the
display reference in seconds.
Return Format
<pos><NL>
<pos> ::= time in seconds from the trigger to the display reference
in NR3 format
See Also
•
"Introduction to :TIMebase Commands"
•
•
•
•
•
NOTE
This command is an alias for the :TIMebase:DELay command.