:SEARch Commands
:SEARch:SERial:LIN:ID
Command Syntax
:SEARch:SERial:LIN:ID <value>
<value> ::= 7-bit integer in decimal, <nondecimal>, or <string>
from 0-63 or 0x00-0x3f
<nondecimal> ::= #Hnn where n ::= {0,..,9 | A,..,F} for hexadecimal
<nondecimal> ::= #Bnn...n where n ::= {0 | 1} for binary
<string> ::= "0xnn" where n ::= {0,..,9 | A,..,F} for hexadecimal
The :SEARch:SERial:LIN:ID command specifies the frame ID value when searching
for LIN events.
Query Syntax
:SEARch:SERial:LIN:ID?
The :SEARch:SERial:LIN:ID? query returns the current frame ID setting.
Return Format
<value><NL>
<value> ::= 7-bit integer in decimal