578
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
23
:MTESt Commands
:MTESt:RMODe:SIGMa
Command Syntax
:MTESt:RMODe:SIGMa <level>
<level> ::= from 0.1 to 9.3 in NR3 format
When the :MTESt:RMODe command is set to SIGMa, the :MTESt:RMODe:SIGMa
command sets the
test sigma
level to which a mask test runs.
Test sigma
is the
best achievable process sigma, assuming no failures. (
Process sigma
is calculated
using the number of failures per test.) The
test sigma
level indirectly specifies the
number of waveforms that must be tested (in order to reach the sigma level).
Query Syntax
:MTESt:RMODe:SIGMa?
The :MTESt:RMODe:SIGMa? query returns the current Sigma level setting.
Return Format
<level><NL>
<level> ::= from 0.1 to 9.3 in NR3 format
See Also
•
"Introduction to :MTESt Commands"
•