:TRIGger Commands
31
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
1129
:TRIGger:PXI:MODE
Command Syntax
:TRIGger:PXI:MODE <mode>
<mode> ::= {OFF | MASTer | SLAVe}
The :TRIGger:PXI:MODE command specifies the PXIe oscilloscope multi-module
triggering mode:
• OFF — This module does not participate in multi-module triggering and is a
standalone module.
• MASTer — This module is the master module.
The :TRIGger commands on the master module define the trigger condition.
After waiting for "armed" signals from each slave module, the master module
outputs the trigger signal on the selected trigger line.
• SLAVe — This module is a slave module.
Slave modules send an "armed" signal on their selected arm like and are
triggered by the selected PXIe trigger line.
It is important to identify the master slot number on each slave
(:TRIGger:PXI:MSLot) so that the proper set of calibration factors can be used
when aligning waveforms. Calibration factors are created and saved using the
KtScopeCal.exe application on the chassis controller.
Among all the oscilloscope modules to be synchronized, one module must be the
master and all other modules must be slaves.
Query Syntax
:TRIGger:PXI:MODE?
The :TRIGger:PXI:MODE? query returns the PXIe oscilloscope's multi-module
triggering mode.
Return Format
<mode><NL>
<mode> ::= {OFF | MAST | SLAV}
See Also
•
":TRIGger:PXI:MALine<n>:ENABle"
•
•
•
•
•